Ufa: diagnostics and repair of 4 Elekta linear accelerators

From June 25 to June 27, 2025, service engineers Pavel Fukalov and Evgeny Smirnov in Ufa, in accordance with the contract, carried out diagnostics and performed repairs on 4 linear accelerators. On Elekta Infinity, we eliminated the table play in the continue direction and eliminated the Gantry Stop error by reworking the potentiometer assembly. On the Elekta Versa HD, a short circuit in the patient table height control circuit has been repaired. Using Elekta Synergy 1, we calibrated the patient table potentiometers; Elekta Synergy 2 identified the cause of the Diode O/L error due to breakdowns in the magnetron.