Diagnostics of MLC Agility in Kaluga

The head of the service department, Konstantin Fukalov, carried out diagnostics of the Agility Head in Kaluga on May 10-11, 2023. The reference reflectors “floated away” - it is necessary to change the lens of the infrared camera in the optical system. The camera and lens have been configured, the image capture system has been calibrated - the device is operational and is “awaiting” the arrival of a new spare part.