Penza: fault diagnosis of two Elekta linear accelerators
From July 6 to July 8, 2026, service engineer Konsantin Lundin in Penza, in accordance with the contract, carried out diagnostics of the Elekta Versa HD linear accelerator, identified a faulty cooling system chiller control controller and thermistor, requiring replacement. He also carried out diagnostics of the Elekta Infinity linear accelerator and identified faulty parts:
- thyratron
- magnetron
- LMDB blade control board and Agility Head cooling fans
- LED Control board
- thyratron control board
- modulator resistor, needs to be replaced.