Chita: diagnostics of the vacuum system of the Elekta linear accelerator
From June 23 to June 27, 2025, service engineer Sergei Chuvakhin in Chita, using an Elekta Synergy linear accelerator, diagnosed the vacuum system using a gel leak detector; no leaks were detected. With the use of a replacement form-vacuum pump, the vacuum pressure was restored to the operating value.